Healing of a coronary artery dissection detected by intravascular ultrasound and optical coherence tomography

Roberto Diletti, MD; Hector M. Garcia-Garcia, MD, PhD; Salvatore Brugaletta, MD; Josep Gomez Lara, MD; Patrick W. Serruys*, MD, PhD, FESC; Evelyn Regar, MD, PhD

A 56-year-old woman, active smoker with no other cardiovascular risk factors was admitted to hospital to undergo elective angioplasty on the proximal circumflex artery. A bioresorbable scaffold BVS 3.0×18mm was implanted. Post-implantation IVUS showed a residual coronary dissection at the distal edge of the scaffold (online Figure). In addition, OCT was performed showing an intimal flap reaching the first frame proximal to the take-off of the side branch distal to the scaffold (Figure1). The dissection was left untreated due to the lack of angiographic evidence of intimal flap and TIMI flow grade 3.1

Figure 1. OCT performed after the procedure revealed a coronary dissection (Panel A). At six months follow-up, there was no evidence of dissection but an area of high reflectivity exactly in the same region according to the position of the side branch used as landmark (Panel B).

After six months, coronary angiogram, IVUS-VH and OCT were repeated showing a complete healing of the dissection, however by OCT there was a clearly defined region of high reflectivity (Figure1). We hypothesise that this increase in reflectivity could be related to a local increase in fibrotic tissue and inflammatory cells2 as a result of the wound healing process and should be considered as an intimal scar. There have been previous reports in the IVUS era describing the fate of edge dissections3. However, this is one of the first observations using OCT. The high sensitivity of this technology allows us to observe the mechanisms of repair of the dissection and could be used in longitudinal studies to observe the evolution over time of residual non-obstructive dissections. The study of these complications may have important clinical implications, in order to define which of these dissections deserve further treatment during the index procedure according to the follow-up findings.

Conflict of interest statement

The authors have no conflicts of interest to declare.

Online data supplement

Figure. After the procedure IVUS clearly identified a dissection at the distal edge of the scaffold but no evidence of dissection could be found at the angiogram (middle panels). At six month follow-up there was no angiographic and no IVUS evidence of dissection and virtual histology did not show any remarkable change in plaque composition (right panels).

Moving image 1. Oct run portraying the dissection post procedure.

Moving image 2. Oct run at six months with no evidence of dissection.

Volume 7 Number 2
Jun 30, 2011
Volume 7 Number 2
View full issue


Key metrics

Suggested by Cory

INTERVENTIONAL FLASHLIGHT

10.4244/EIJ-D-17-00192 Oct 20, 2017
Long-term outcome of a spontaneous coronary artery dissection treated with a bioresorbable scaffold
free

IMAGE IN CARDIOLOGY

10.4244/EIJV10I11A225 Mar 20, 2015
The “plastic healing concept”: implantation of bioabsorbable scaffolds in spontaneous coronary artery dissection
Al Mamary A et al
free

SHORT REPORT

10.4244/EIJ-D-18-00519 Jan 18, 2019
Long-term follow-up of spontaneous coronary artery dissection treated with bioresorbable scaffolds
Macaya F et al
free

Image – Interventional flashlight

10.4244/EIJ-D-19-00063 Jan 20, 2021
Late structural discontinuity after bioresorbable vascular scaffold implantation in patients with in-stent restenosis
Cuesta J et al
free

IMAGE IN CARDIOLOGY

10.4244/EIJV12I7A142 Sep 18, 2016
Plaque protrusion compromising bioresorbable coronary scaffold patency
Floré V et al
free
Trending articles
69.996

10.4244/EIJV13I12A217 Dec 8, 2017
Swimming against the tide: insights from the ORBITA trial
Al-Lamee R and Francis D
free
59.65

State-of-the-Art

10.4244/EIJ-D-24-00066 Apr 21, 2025
Management of complications after valvular interventions
Bansal A et al
free
57.6

State-of-the-Art

10.4244/EIJ-D-24-00386 Feb 3, 2025
Mechanical circulatory support for complex, high-risk percutaneous coronary intervention
Ferro E et al
free
38.75

State-of-the-Art

10.4244/EIJ-D-23-00912 Oct 7, 2024
Optical coherence tomography to guide percutaneous coronary intervention
Almajid F et al
free
15.85

State-of-the-Art

10.4244/EIJ-D-23-01050 Jul 15, 2024
Durability of transcatheter aortic valve implantation
Ternacle J et al
free
X

The Official Journal of EuroPCR and the European Association of Percutaneous Cardiovascular Interventions (EAPCI)

EuroPCR EAPCI
PCR ESC
Impact factor: 7.6
2023 Journal Citation Reports®
Science Edition (Clarivate Analytics, 2024)
Online ISSN 1969-6213 - Print ISSN 1774-024X
© 2005-2025 Europa Group - All rights reserved